Skip to main content
. 2014 Feb 24;41(3):031908. doi: 10.1118/1.4865788

TABLE V.

Imaging dose data for the Elekta XVI system. Measurements were acquired for the full rotation Pelvis M20 protocol, full rotation Chest M20 protocol, and partial rotation Prostate Seed S10 protocol.

Elekta XVI
    Pelvis M20 Chest M20 Prostate Seed S10
    Additional 4 cm   Additional 4 cm   Additional 4 cm
Location Nominal thickness (cGy) diameter (cGy) Nominal thickness (cGy) diameter (cGy) Nominal thickness (cGy) diameter (cGy)
Interior Anterior 2.32 ± 0.12 1.85 ± 0.07 2.49 ± 0.13 2.00 ± 0.10 0.45 ± 0.03 0.33 ± 0.03
  Posterior 2.34 ± 0.12 1.76 ± 0.06 2.44 ± 0.12 1.80 ± 0.09 0.29 ± 0.03 0.17 ± 0.02
  Left 1.65 ± 0.08 1.40 ± 0.05 1.63 ± 0.08 1.43 ± 0.07 0.48 ± 0.03 0.34 ± 0.03
  Right 1.92 ± 0.10 1.54 ± 0.06 1.92 ± 0.10 1.56 ± 0.08 0.21 ± 0.02 0.14 ± 0.01
  Center 2.14 ± 0.11 1.56 ± 0.06 2.15 ± 0.11 1.68 ± 0.08 0.33 ± 0.03 0.21 ± 0.02
Surface Anterior 2.73 ± 0.14 2.21 ± 0.11 2.41 ± 0.12 2.22 ± 0.11 0.66 ± 0.03 0.65 ± 0.03
  Posterior 2.57 ± 0.13 2.17 ± 0.11 2.74 ± 0.14 2.15 ± 0.11 0.33 ± 0.03 0.30 ± 0.03
  Left 1.50 ± 0.08 1.49 ± 0.08 1.64 ± 0.08 1.48 ± 0.07 0.55 ± 0.03 0.50 ± 0.03
  Right 1.71 ± 0.09 1.72 ± 0.09 1.66 ± 0.08 1.80 ± 0.07 0.02 ± 0.01 0.01 ± 0.01