TABLE I.

Determination of D0 for photon beams using 10-cm normalization depth. For a nonisocentric system with 1 MU defined to deliver 1 cGy at 100 cm SSD and depth = dm, the dose/MU is calculated at the suggested normalization conditions of d = 10 cm, 100 cm SAD, using depth of dm and TMR data taken from Ref. 57. The suggested D0 maintains the definition 1 MU under a new, 10-cm normalization condition.

Energy d m TMR   Dose/MU Suggested D0
(MV) (cm) (10, 10×10) 100+dm1002 (cGy/MU) (cGy/MU)
4 1 0.738 1.020 0.753 0.750
5 1.25 0.759 1.025 0.778 0.780
6 1.5 0.786 1.030 0.810 0.810
8 2 0.820 1.040 0.853 0.850
10 2.3 0.839 1.047 0.878 0.880
12 2.6 0.858 1.053 0.903 0.900
15 2.9 0.877 1.059 0.929 0.930
18 3.2 0.896 1.065 0.954 0.950
21 3.5 0.914 1.071 0.979 0.980
25 3.8 0.933 1.077 1.005 1.000