Electron output factors. Electron output factors versus field size for various applicators for a 9 MeV beam on a Varian Clinac 2100C (SSD = 100 cm).

Field size Applicator size (cm2)
(cm2) 10 × 10 15 × 15 20 × 20 25 × 25
2 × 2 0.843 0.846 0.824 0.812
3 × 3 0.908 0.908 0.891 0.871
4 × 4 0.963 0.954 0.936 0.911
5 × 5 0.991 0.982 0.962 0.936
6 × 6 1.003 0.997 0.976 0.951
7 × 7 1.005 1.004 0.983 0.960
8 × 8 1.003 1.006 0.986 0.964
10 × 10 1.000 1.003 0.985 0.965
12 × 12   0.997 0.980 0.962
15 × 15   0.992 0.976 0.955
20 × 20     0.981 0.950
25 × 25       0.955