Skip to main content
. 2014 Feb 26;41(3):031501. doi: 10.1118/1.4864244

TABLE VIII.

Electron SSDeff table. Table of SSDeff averaged over applicator size for a Varian Clinac 2100C. For this table, SSDeff is a function of insert size and energy only.

Insert size Energy (MeV)
(cm2) 6 9 12 16 20
4 46.2 61.1 72.5 76.4 76.6
6 62.2 74.7 80.2 81.8 80.6
8 77.6 83.8 83.2 83.6 82.7
10 82.9 85.9 86.8 85.5 83.8
15 90.7 90.9 90.1 90.0 89.7
20 90.0 91.8 91.4 90.9 92.0
25 90.7 91.9 91.0 92.5 93.3