TABLE II.

Effects of systematic longitudinal source positioning errors on dose to the HR-CTV, and organs at risk for 45° azimuthal emission angle. Only the organ-at-risk D2cc values for which the D2cc tolerance was violated following the positioning errors were included. Organ-at-risk tolerance D2cc-values for sigmoid colon, rectum, and bladder were 75, 75, and 90 Gy3, none of which were violated when there is no positioning error.

Patient Longitudinal positioning error (mm) HR-CTV D90 change (%) Sigmoid colon D2cc (Gy3) Rectum D2cc (Gy3) Bladder D2cc (Gy3)
1 −1 −2.15 76.18
+1 −2.07 93.26
2 −1 −6.31 97.89
+1 −6.17 98.24
3 −1 −3.54 75.89
+1 −3.4 75.97
4 −1 −15.67 78.71
+1 −5.18 102.24
5 −1 −6 75.62
+1 −4.96 105.8