TABLE V.
Summary statistics for Fig. 5. Paired t-test and F-test p-values were calculated by testing the clinical (repeatability) distribution against each reduced-dose reproducibility distribution.
Dose level | N | Mean (%) | SD (%) | Paired t-test p-value | F-test p-value |
---|---|---|---|---|---|
Clinical (repeatability) | 17 | 1.1 | 5.5 | ||
25% vs clinical | 17 | −4.5 | 6.3 | 0.04a | 0.22 |
10% vs clinical | 17 | −5.6 | 9.9 | 0.03a | 0.0002a |
3% vs clinical | 17 | −1.7 | 6.8 | 0.53 | 0.11 |
The p-value was statistically-significant.