Skip to main content
. 2016 Summer;15(3):379–391.

Table 6.

Analytical figures of merit of the spectrophotometric method and PLS1, PCR and HLA regressions.



THEO


MKST


LORA

Parameters PLS1 PCR HLA PLS1 PCR HLA PLS1 PCR HLA
SEN 0.119 0.127 0.018 0.062 0.069 0.040 0.081 0.083 0.022
SEL 0.464 0.498 0.072 0.178 0.198 0.114 0.342 0.350 0.094
LOD 0.03 0.03 0.18 0.06 0.05 0.09 0.05 0.05 0.10
γ −1 0.010 0.010 0.059 0.020 0.016 0.030 0.016 0.016 0.033

The unit of SEN is (AU. L.mg–1) that AU is absorbance unit, unit of γ (L mg–1) and unit of γ–1 and LOD (mg L–1).