TABLE 5.

Sensitivity, specificity, and agreement of three LD WB tests and the C6 ELISA for the BBI and CDC performance panelsa

Ig and test Relative sensitivity (%) Relative specificity (%) Overall agreement (%)
IgG
    Marblot 85 (89/83) 97 (100/96) 91 (93/90)
    Virablot 85 (89/83) 90 (100/88) 88 (93/86)
    BBI WB 81 (89/78) 97 (100/96) 89 (93/88)
IgM
    Marblot 62 (43/68) 97 (88/100) 81 (67/86)
    Virablot 85 (86/84) 77 (88/74) 81 (87/79)
    BBI WB 77 (71/79) 100 (100/100) 89 (87/90)
IgG/IgM
    C6 ELISA 100 (100/100) 73 (100/69) 93 (100/90)
a

Data are presented as BBI + CDC (BBI/CDC), where BBI indicates percentages for the BBI mixed-titer performance panel and CDC indicates percentages for the CDC LD evaluation panel.