Table 3. Results of accuracy, precision and recovery of method validation using quality control samples.

Analytes QC (ng/mL) Intra-day (n = 5) Inter-day (n = 5) Recovery (%)
Accuracy (%) Precision (%) Accuracy (%) Precision (%)
Serum
 24S-OHC 5 96.0 4.7 97.7 10.4 89.8
10 95.0 4.1 99.7 4.5 88.1
100 106.0 0.5 100.9 3.0 85.8
200 99.1 0.9 100.1 1.2 82.7
 25-OHC 0.5 109.7 8.2 93.3 14.7 90.7
1 98.6 5.5 92.5 7.2 89.7
10 103.0 0.5 105.1 4.1 87.5
20 94.9 0.9 97.4 2.2 83.7
 27-OHC 5 104.5 8.2 107.3 7.5 85.8
10 103.8 5.5 107.4 4.2 87.5
100 99.7 0.5 96.0 1.9 88.7
200 99.7 0.8 101.1 1.0 84.0
CSF
 24S-OHC 0.1 97.1 7.6 96.2 4.5 83.7
0.5 101.6 5.3 108.1 2.3 95.8
2 98.7 1.4 98.0 1.8 97.2
5 99.7 1.6 98.5 0.8 92.6
 25-OHC 0.03 112.9 7.3 119.2 4.5 68.4
0.1 107.8 4.0 109.2 3.9 98.5
0.5 99.6 1.3 101.0 2.3 99.6
1 107.5 1.2 110.3 3.2 93.2
 27-OHC 0.1 115.8 7.7 108.3 3.5 88.2
0.5 101.3 1.8 100.8 1.6 105.0
2 97.1 2.3 94.5 2.4 103.2
5 99.8 1.2 96.3 2.6 97.3