Skip to main content
. 2016 Dec 3;10:210–214. doi: 10.1016/j.dib.2016.11.084

Table 2.

Data for 4 boards on each sides of the phantom.

Reconstruction Initial dose −10% dose −20% dose −30% dose
eAIDR FIRST FIRST FIRST
Tube current (mA) 200 180 160 140
Standard deviation (Hounsfield units) 16.67 16.44 16.37 16.62
Subjective scores
 4 mm phantom 3 4 4 5
 3 mm phantom 3 4 4 4
 2 mm phantom 3 4 4 4