Table 3. Comparison of the analytical performance of the proposed MSPE-UPLC-MS/MS method with other ones for determination of PFCs in surface water sample.

Analytical methods Adsorbents Sorbent amount (mg) Sample volume (mL) Enrichment factor Extraction time (min) Analytical timea (min) Linearity ranges (ng L−1) LODs (ng L−1) Spike recoveries (%) RSDs (%)
MSPE-UPLC-MS/MS25 Fe3O4@SiO2-FBC 40 100 100 12 3.5 0.25–25 0.01–0.06 89.34–111.32 <9.9
MSPE-HPLC-MS/MS19 Fe3O4-C18@Chitosan 100 500 1000 20 6.5 0.5–50 0.05–0.24 56–112 2–8
MSPE-HPLC-MS/MS17 Fe3O4-CTAB 75 800 1600 10 6.5 0.5–50 0.03–0.88 66–123 1–8
SPE-HPLC-MS/MS13 Oasis-Wax 150 1000 1000 >300 >10 0.2–50 <0.1–0.6b 61–118 <18
SPE-HPLC-MS/MS15 Oasis-HLB 200 500 500 >160 >20 n. r.c <0.2–2.0b 77–137 n. r.c
This work Fe3O4@SiO2-NH2&F13 50 500 1000 30 8 0.5–100 0.029–0.099 90.05–106.67 3.1–12.6

aConfirmed by the time of chromatogram peak appearance of the last target analyte.

bLOQs. The LODs were not given in the References.

cNot reported.