Table 1.

Prototype assay calibration performance summary using Quanterix standard for assay evaluation and clinical test batches

Concentrationa (pg/ml)
Theoretical concentration 0.023 0.069 0.206 0.617 1.85 5.6 16.7 50.0
 Run 1 0.096 0.199 0.616 1.85 5.56 16.9 49.9
 Run 2 −0.011 0.066 0.224 0.614 1.77 5.67 17.2 49.7
 Run 3 −0.032 0.135 0.281 0.613 1.78 5.56 17.4 49.5
 Run 4 −0.012 0.080 0.201 0.615 1.87 5.48 16.9 49.9
 Run 5 −0.074 0.103 0.218 0.602 1.85 5.54 16.9 49.9
 Run 6 0.061 0.088 0.194 0.620 1.85 5.48 17.0 49.8
 Run 7 0.066 0.202 0.627 1.82 5.56 16.8 49.9
 Run 8 0.234 0.323 0.525 2.11 5.23 17.5 49.8
 Run 9 0.214 0.240 0.583 1.82 5.43 17.8 49.2
N (total number of runs) 5 9 9 9 9 9 9 9
Inter-assay mean −0.014 0.120 0.231 0.602 1.86 5.5 17.2 49.7
SD 0.049 0.063 0.044 0.031 0.099 0.121 0.350 0.237
CV% −362 52.1 19.0 5.2 5.3 2.2 2.0 0.5
Inter-assay RE, % −159 74.2 12.4 −2.5 0.4 −1.8 2.7 −0.5

CV% Coefficient of variation percentage, RE Relative error

In runs 6–8, calibration curves prepared using both Quanterix and Fujirebio standards were included

aConcentrations were interpolated using the calibrator average enzymes per bead values against the fitted cubic curves