Table 2.

Prototype assay calibration performance summary using Fujirebio standard for assay evaluation and clinical test batches

Concentrationa (pg/ml)
Theoretical concentration 0.023 0.069 0.206 0.617 1.85 5.6 16.7 50.0
 Run 6 −0.006 0.097 0.262 0.584 1.83 5.65 16.6 50.1
 Run 7 0.112 0.365 0.514 1.94 5.36 17.1 49.8
 Run 8 0.260 0.283 0.388 1.98 5.79 16.0 50.5
N (total number of runs) 1 3 3 3 3 3 3 3
Inter-assay mean −0.006 0.156 0.303 0.495 1.92 5.6 16.6 50.1
SD 0.090 0.054 0.100 0.077 0.220 0.508 0.365
CV% 57.9 17.9 20.1 4.0 3.9 3.1 0.7
Inter-assay RE, % −125 128 47.5 −19.8 3.6 0.8 −0.7 0.3

CV% Coefficient of variation percentage, RE Relative error

In runs 6–8, calibration curves prepared using both Quanterix and Fujirebio standards were included

aConcentrations were interpolated using the calibrator average enzymes per bead values against the fitted cubic curves