. 2016 Dec 15;8:58. doi: 10.1186/s13195-016-0225-7

Table 3.

Summary of average enzymes per bead for prototype assay calibration using Quanterix standard for assay evaluation and clinical test batches

AEB
Theoretical concentration, pg/ml 0.023 0.069 0.206 0.617 1.85 5.6 16.7 50.0
 Run 1 0.006 0.008 0.010 0.023 0.112 0.794 5.951 27.615
 Run 2 0.005 0.006 0.008 0.018 0.086 0.692 5.214 23.725
 Run 3 0.005 0.007 0.009 0.018 0.089 0.703 5.774 29.431
 Run 4 0.006 0.006 0.008 0.020 0.102 0.720 5.665 28.302
 Run 5 0.005 0.006 0.008 0.018 0.096 0.720 5.603 28.131
 Run 6 0.006 0.007 0.009 0.019 0.089 0.624 5.082 27.607
 Run 7 0.011 0.012 0.014 0.025 0.091 0.618 4.674 25.809
 Run 8 0.007 0.009 0.011 0.015 0.107 0.580 5.454 24.691
 Run 9 0.007 0.009 0.010 0.018 0.084 0.608 5.409 24.587
N (total number of runs) 9 9 9 9 9 9 9 9
Inter-assay mean 0.006 0.008 0.010 0.019 0.095 0.673 5.425 26.655
SD 0.00 0.00 0.00 0.00 0.01 0.07 0.39 2.00
CV% 26.4 24.8 19.1 15.4 10.4 10.3 7.2 7.5

AEB Average enzyme per bead, CV% Coefficient of variation percentage

In runs 6–8, calibration curves prepared using both Quanterix and Fujirebio standards were included