Table 4.

Summary of average enzymes per bead for prototype assay calibration using Fujirebio standard for assay evaluation and clinical test batches

AEB
Theoretical concentration, pg/ml 0.023 0.069 0.206 0.617 1.85 5.6 16.7 50.0
 Run 6 0.006 0.007 0.008 0.012 0.031 0.164 1.105 8.401
 Run 7 0.011 0.012 0.014 0.015 0.036 0.142 1.083 8.115
 Run 8 0.011 0.012 0.013 0.013 0.033 0.149 0.931 8.084
N (total number of runs) 3 3 3 3 3 3 3 3
Inter-assay mean 0.009 0.010 0.012 0.013 0.033 0.151 1.040 8.200
SD 0.003 0.003 0.003 0.002 0.002 0.011 0.095 0.175
CV% 28.8 29.1 25.0 13.8 6.6 7.3 9.1 2.1

AEB Average enzyme per bead, CV% Coefficient of variation percentage

In runs 6–8, calibration curves prepared using both Quanterix and Fujirebio standards were included