Skip to main content
. 2016 Dec 15;8:58. doi: 10.1186/s13195-016-0225-7

Table 5.

Estimated amyloid-β 1–42 peptide content of nominal 0.25-ng masses of calibration standard

Calibration standard Theoretical amount loaded (ng) Mean amount measured (ng) Relative error (%) SEM
Quanterixa 0.250 0.367 +47 8.1
Fujirebioa 0.250 0.123 −51 7.3
Lillyb 0.250 0.230 −8 15.0

a n = 3