Table 7.

Summaries of intra- and inter-assay precision for repeated testing of three quality control samples and one plasma pool

1–42 (pg/ml)
Sample QC1 QC2 QC3 Plasma poola
Intra-assay means (n = 4) 20.5 4.89 0.696 5.32
20.0 4.87 0.628 5.00
19.6 4.76 0.745 5.32
Intra-assay CV% 2 2 1 1
4 1 7 1
0.4 0.1 8 3
Inter-assay means (n = 3) 20.0 4.84 0.689 5.20
Inter-assay CV% 2 2 8 4

amyloid-β, CV% Coefficient of variation percentage, QC Quality control

a n = 18 for plasma pool; the reported Aβ1–42 levels in plasma pool were corrected for four-fold sample dilution