TABLE 2.

Structural properties for each mesh obtained from load-elongation curves to 5N

Mesh (n = 5 each) Elongation, mm±SD Low stiffness, N/mm±SD High stiffness, N/mm±SD
Gynemesh 12.78 ± 0.00 0.30 ± 0.01 0.45 ± 0.02
Restorelle 0-degree 16.00 ± 0.03 0.16 ± 0.07 0.41 ± 0.06
Restorelle 5-degree 17.42 ± 0.03 0.23 ± 0.03 0.32 ± 0.04
Restorelle 45-degree 26.70 ± 0.06 0.02 ± 0.01 0.42 ± 0.03
Alyte Vaginal 12.67 ± 0.01 0.12 ± 0.01 0.51 ± 0.02
Alyte Stem 9.76 ± 0.01 0.24 ± 0.04 0.55 ± 0.02
UltraPro 14.41 ± 0.01 0.08 ± 0.01 0.49 ± 0.02
UltraProOpp 59.25 ± 0.02 0.01 ± 0.01 0.27 ± 0.02
Overall probability value .001 .000 .029

Elongation values are those that resulted from 5N of force being applied to the mesh; low and high stiffness values are the minimum and maximum slopes of the load-elongation curve, respectively. To minimize the complexity of this Table, relevant probability values for individual comparisons are reported in the text.