Table 1.

Summary of benchmark credentialing attempts, reasons for not passing, and planning techniques used in the 59 passing benchmarks.

No. Percentage
Number of attempts at passing benchmark (n = 63)
1 32 51%
2 24 38%
3 3 5%
No further attempts 4 6%
Reasons for not passing benchmark at first attempt (n = 31)
Separate plans did not meet OAR criteria 7 23%
PTV coverage unacceptably high or low 7 23%
Stomach dose constraints exceeded 3 10%
Spinal dose constraints exceeded 7 23%
Conformity does not meet SBRT guideline 5 16%
Other 2 6%
Planning technique used to create passing benchmark (n = 59)
VMAT 46 78%
IMRT 8 14%
3DCRT 1 2%
Robotic Radiosurgery 4 7%
Number of plans and isocenters utilized (n = 59)
1 plan/1 isocenter 39 66%
1 plan/2 isocenters 14 24%
2 plans 2 3%
Robotic Radiosurgery 4 7%
Structures prioritized (n = 59)
PTV 44 75%
Stomach 15 25 %