Skip to main content
. 2016 Dec 1;18(12):800–805. doi: 10.1089/dia.2016.0279

Table 3.

Receiver Operating Characteristic Curve Analysis Was Used to Evaluate the Diagnostic Performance of NerveCheck for Sensory Loss Against Sural Nerve Action Potential, Sural Nerve Conduction Velocity, Corneal Nerve Fiber Density (No./mm2), Intraepidermal Nerve Fiber Density (No./mm), and McGill Questionnaire for Symptoms of Painful Neuropathy

  VPT, AUC% (95% CI), P CPT, AUC% (95% CI), P WPT, AUC% (95% CI), P
SNAP 82 (72–93), <0.0001 70 (56–84), 0.005 78 (65–90), 0.0002
SNCV 84 (75–94), <0.0001 71 (58–84), 0.002 75 (63–87), 0.0003
CNFD 76 (66–87), 0.0002 78 (66–91), <0.0001 71 (59–84), 0.002
IENFD 60 (42–79), 0.2 70 (54–87), 0.01 69 (52–85), 0.03
McGill (pain) 70 (57–83), 0.006 63 (48–78), 0.07 65 (50–80), 0.04

AUC, area under the curve; CI, confidence interval.