TABLE III.

The logic implemented to alert clinicians if maternal and fetal heart rate, maternal oxygen saturation and maternal blood pressure readings are abnormal. Readings that require urgent attention trigger an automated call at the end of the encounter, all other measurements are included in the encounter summary SMS to inform medical staff. Blood pressure readings are currently transmitted as photo and are therefore addressed when reviewed by medical staff after data upload; optical character recognition is being implemented for the next iteration to convert the image into numeric values automatically triggering the appropriate alert. There are currently no Bluetooth enabled blood pressure monitors approved for use in pregnancy (for the detection of pre-eclampsia).

Measurement Logic Detection trigger Indication
Heart rate [bpm] Maternal HR ≤ 60 SMS alert Maternal bradycardia
Maternal HR ≥ 100 SMS alert Maternal tachycardia
Fetal HR ≤ 120 Call Fetal bradycardia [27]
Fetal HR ≥ 160 Call Fetal tachycardia [27]

Oxygen saturation [%] Oxygen ≤ 90 Call Maternal hypoxemia [28]

Blood pressure [mmHg] Systolic ≤ 70 Current: Addressed at data review Low blood pressure
Future: Call
Systolic ≥ 140, diastolic ≥ 90 Current: Addressed at data review Mild hypertension [29]
Future: SMS alert
Systolic ≥ 155, diastolic ≥ 105 Current: Addressed at data review Severe hypertension [29]
Future: Call