Median Saturations and Proportion of Time Spent <85% (Lower Tail) and >95% (Higher Tail) Based on Pulse Oximeter Algorithm (Original or Revised) While on Supplemental Oxygen

Study Target Spo2 Arm Median Spo2, % Tail, Proportion of Time Spent at This Spo2 Level, % Original, % Revised, %
Original Revised
COTa 85%–89% (low) 91 91 <85 20.2 18.5
>95 17.5 15.2
91%–95% (high) 93 93 <85 9.1 8.2
>95 22.4 21.2
BOOST-II UKb 85%–89% (low) 91 90 <85 25.7 22.1
>95 16.1 13.9
91%–95% (high) 92 93 <85 15.0 12.3
>95 18.7 20.4
BOOST-II Australiab 85%–89% (low) 90 89 <85 27.4 24.1
>95 11.1 8.1
91%–95% (high) 93 92 <85 13.5 10.8
>95 18.6 16.4
BOOST-II New Zealandb 85%–89% (low) 91 <85 21.1 Only oximeters with the original algorithm were used in these studies.
>95 15.5
91%–95% (high) 93 <85 10.8
>95 22.2
SUPPORTb 85%–89% (low) 90c <85 10.9
>95 33.0
91%–95% (high) 93c <85 16.0
>95 26.7
a

All study days with >12 hours of supplemental O2.

b

Time when the infant was receiving oxygen.

c

Estimate based on visual inspection of the graphs.