Table 2.

Parameters in EOS micro-dose and standard digital radiography

EOS micro-dose Digital radiography
FSD (cm) 86.43 (4.56) 163.3 (0.94)
Beam width (cm) 44.22 (1.27) 34.2 (2.51)
Beam height (cm) 76.69 (4.47) 32.9 (3.04) – in 3 sections
Projection angle (degree) 90o (PA) 90o (PA)
X-ray tube potential (kv) 60.7 (1.83) 78.2 (5.9)
X-ray current (mA) 80.8 (3.96) 402.3 (56.0)
X-ray tube anode angle (degree) 7o 12o
Filter Copper (Cu) Aluminum (Al)
Filter thickness (mm) 0.1 2.7
Scanning Time (sec/msec) 7.72 (0.65) 12.09 (3.92) Sternal notch
33.03 (13.72) Nipple line
53.14 (18.04) Symphysis pubis
Monte Carlo simulation parameters: Maximum energy 150 keV 150 keV
Monte Carlo simulation parameters: Number of photons 20000 20000