Table:
Laboratory | RCPAQAP Material Slope | Common Calibrator Material Slope | Unknown Fresh frozen serum Slope |
---|---|---|---|
1* | 0.986 | 0.985 | 1.0399 |
2* | 0.9784 | 0.9667 | 0.9989 |
3** | 0.8949 | 0.9369 | 0.8915 |
4** | 0.9196 | 0.9706 | 0.9406 |
5* | 0.9961 | 1.0268 | 0.992 |
6*** | 0.9252 | 1.0834 | 0.9539 |
7* | 1.0251 | 1.019 | 1.0286 |
8* | 1 | 1 | 0.9621 |
Summary of slopes | |||
Minimum slope | 0.895 | 0.937 | 0.892 |
Maximum slope | 1.025 | 1.083 | 1.040 |
Mean slope | 0.966 | 0.999 | 0.976 |
Median slope | 0.982 | 0.993 | 0.977 |
SD of slope | 0.046 | 0.045 | 0.049 |
Minus 2SD of slope | 0.873 | 0.908 | 0.878 |
Plus 2SD of slope | 1.058 | 1.089 | 1.074 |
Potential overall difference (95% range) between slopes | 19% | 18% | 20% |
R2 was >0.99 for all materials
R2 was >0.99 for Common Calibrator & Fresh frozen serum, but < 0.99 for the RCPAQAP material assessment (R2= 0.988 for lab 3) and (R2 = 0.9771 for lab 4)
R2 was >0.99 for the RCPAQAP material, but < 0.99 for the Common Calibrator (R2 = 0.9891) and the Fresh frozen serum (R2 = 0.9883)
Note: Laboratory 8 was the reference laboratory, i.e. NMIA LC-MS/MS method.