Skip to main content
. 2017 Jan 5;7:40090. doi: 10.1038/srep40090

Table 1. Precision data for LTIA based on 3 monoclonal antibodies.

Concentration (ng/mL) Within-run (n = 20)
Total (n = 20)
Mean S.D. CV (%) Mean S.D. CV (%)
25 25.421 0.733 2.885 25.215 1.824 7.234
100 100.671 1.524 1.514 100.721 6.539 4.507
150 152.029 4.41 2.901 151.989 5.520 3.632