Skip to main content
. 2016 Jan 22;77(1):34–43. doi: 10.1002/mrm.26103

Table 4.

Metabolite COVs for Voxels Containing Spectra That Meet “Acceptable” Water Line Width Criteria

0.06‐0.1ppm
TE = 35 ms TE = 80 ms TE = 144 ms Echo time comparison
Line width COV % COV % COV % ANOVA
Metabolite Si.BL Ex.BL Si.BL Ex.BL Si.BL Ex.BL Si.BL Ex.BL
tNAA††¥¥ 11.94 8.41 6.77 5.92 14.73 8.27 <0.00001 <0.0001
tCr 9.78 9.74 8.12 8.33 14.02 12.73 <0.00001 <0.001
tCho 9.75 10.47 9.00 8.98 10.89 11.03 0.2815 0.1056
Glu† 23.06 18.63 24.45 24.68 62.45 59.21 <0.00001 <0.00001
Gln 83.65 85.64 61.78 65.36 48.18 54.81 <0.00001 <0.00001
Glx† 29.56 23.99 24.41 24.54 23.34 26.99 0.01 0.3328
Ins 14.85 15.06 19.49 20.47 47.85 45.72 <0.00001 <0.00001
Glth† 63.00 102.50 28.71 28.53 92.26 88.03 <0.00001 <0.00001
Tau†† 128.16 96.08 73.74 74.43 91.93 86.37 <0.00001 <0.01

Metabolite COVs for TE = 35, 80, and 144 ms for voxels containing spectra that meet the “acceptable” water line width criteria of between 0.06 and 0.1ppm showing significant (P < 0.05) differences from an unpaired t‐test between Si.BL and Ex.BL for TE = 35 ms (†), TE = 80 ms (‡), and TE = 144 ms (¥). Significances of P < 0.01 are denoted as: TE = 35 ms (††), TE = 80 ms (‡‡), and TE = 144 ms (¥¥).