Skip to main content
. 2015 Sep 18;105(1):211–221. doi: 10.1002/jbm.b.33525

Figure 6.

Figure 6

Measured flow rate versus pressure of the InnFocus MicroShunt, which is 8.5 mm long with a lumen diameter of 70 μm. It requires approximately 5 mm Hg pressure to initiate flow through the MicroShunt.