Skip to main content
. Author manuscript; available in PMC: 2017 Jan 6.
Published in final edited form as: Metrologia. 2016;53(Technical Suppl):03005. doi: 10.1088/0026-1394/53/1A/03005

Table 10.

Degree of equivalence between TDP/FP realized by NIST and the KCRV, DNIST/KCRV, and its expanded uncertainty (k = 2), U(DNIST/KCRV), at TDP/FP values of +20 °C, 1 °C, and −10 °C, as given by Tables 7.3 and 7.4 in [1].

Nominal TDP/FP (°C) DNIST/KCRV (°C) U(DNIST/KCRV) (°C)
20 −0.006 0.050
1 −0.011 0.060
−10 −0.039 0.043