Table 5.

Uncertainty elements and their standard uncertainty values for the LACOMET generator, for the four nominal values of TDP/FP.

Uncertainty for LACOMET generator : TDP = +20 °C TDP = +1 °C TFP = −10 °C TFP = −30°C
Saturation temperature (Thermometer and Saturator)
Calibration uncertainty (sensor and indicator unit) °C 0.0050 0.0050 0.0050 0.0050
Long-term stability (sensor and indicator) °C 0.0050 0.0050 0.0050 0.0050
Self-heating and residual heat fluxes (sensor) °C 0.0060 0.0060 0.0060 0.0060
Resolution and accuracy or linearity (indicator unit) °C 0.0005 0.0005 0.0005 0.0005
Temperature homogeneity °C 0.0100 0.0100 0.0100 0.0100
Temperature stability °C 0.0060 0.0060 0.0060 0.0100
Saturation pressure (Pressure gauge)
Calibration uncertainty (sensor and indicator unit) Pa 2.4 2.4 2.4 2.4
Long-term stability (sensor and indicator) Pa 1 1 1 1
Resolution and accuracy or linearity (indicator unit) Pa 1 1 1 1
Pressure differences in the saturator cell Pa 20 20 20 20
Stability of the pressure Pa 10 10 10 10
Gas pressure at the generator outlet (Pressure gauge)
Calibration uncertainty (sensor and indicator unit) Pa 10 10 10 10
Long-term stability (sensor and indicator) Pa 20 20 20 20
Resolution (indicator unit) Pa 20 20 20 20
Stability of the pressure Pa 10 10 10 10
Effect of the tubing between the saturator and the pressure gauge Pa 50 50 50 50
Saturation efficiency °C 0.0080 0.0080 0.0080 0.0080
Uncertainty due to formulae/calculations
Saturation vapor pressure formula(e) °C 0 0 0 0
Water vapor enhancement formula(e) °C 0 0 0 0
Other
Pressure drop between point of realization and measuring instrument Pa 25 25 25 25
Water contaminations °C 0.0010 0.0010 0.0010 0.0010