Skip to main content
. Author manuscript; available in PMC: 2017 Jan 6.
Published in final edited form as: Metrologia. 2016;53(Technical Suppl):03005. doi: 10.1088/0026-1394/53/1A/03005

Table 7.

Standard uncertainty of the determinations of ΔTDP/FP for NIST and LACOMET. The column headings are described in the text.

Nominal TDP/FP (°C) Participating Institute uATDP/FP) (°C) u(TDP/FPg) (°C) ucTDP/FP) (°C)
20 LACOMET 0.003 0.024 0.024
20 NIST 0.003 0.006 0.007
1 LACOMET 0.002 0.023 0.023
1 NIST 0.009 0.010 0.013
−10 LACOMET 0.004 0.022 0.022
−10 NIST 0.009 0.008 0.012
−30 LACOMET 0.003 0.023 0.023
−30 NIST 0.019 0.008 0.021