Standard uncertainty of the determinations of ΔTDP/FP for NIST and LACOMET. The column headings are described in the text.
| Nominal TDP/FP (°C) | Participating Institute | uA (ΔTDP/FP) (°C) | (°C) | uc(ΔTDP/FP) (°C) |
|---|---|---|---|---|
| 20 | LACOMET | 0.003 | 0.024 | 0.024 |
| 20 | NIST | 0.003 | 0.006 | 0.007 |
| 1 | LACOMET | 0.002 | 0.023 | 0.023 |
| 1 | NIST | 0.009 | 0.010 | 0.013 |
| −10 | LACOMET | 0.004 | 0.022 | 0.022 |
| −10 | NIST | 0.009 | 0.008 | 0.012 |
| −30 | LACOMET | 0.003 | 0.023 | 0.023 |
| −30 | NIST | 0.019 | 0.008 | 0.021 |