. 2016 Sep 24;311(1):409–418. doi: 10.1007/s10967-016-5036-6

Table 3.

Correction factors as used

Country Dead time Gamma self-absorption and geometry TCC
Egypt MoO3 1.0015 1.0840 1.0000
1 g 1.0079 1.0720 1.0000
5 g 1.0020 1.0850 1.0000
Morocco MoO3 1.015 1.007
1 g 1.022 1.006
5 g 1.105 1.002
Malaysia MoO3 1 1 1
1 g 1 1 1
5 g 1 1 1
Vietnam MoO3 1.28 (140 keV)
1 g 1.36 (140 keV)
5 g 1.36 (140 keV)
Hungary MoO3 ZDT 1.0406 1
1 g ZDT 1.0020 1
5 g ZDT 1.0050 1
Kazachstan MoO3 1.0042 1
1 g 1.0063 1.014
5 g 1.0383 1.015
Netherlands MoO3 1.0846 1.0638 1.0050
1 g 1.0050 1.0395 1.0050
5 g 1.0299 1.0428 1.0050
Portugal MoO3 LT preset 1.0860
1 g LT preset 1.0360
5 g LT preset 1.0360
Romania MoO3 Genie 1 1
1 g Genie 1 1
5 g Genie 1 1
Slovenia MoO3 ZDT 1.1125 1.0078
1 g ZDT 1.0262 1.0087
5 g ZDT 1.0208 1.0089
Ukraine MoO3 1.0215 1.031
1 g 1.0194 1.031
5 g 1.0846 1.031
Chile MoO3 1.0000 1.0000 1.0000
1 g 1.0000 0.8438 1.0000
5 g 1.0000 0.9694 1.0000
Mexico MoO3 1.0054
1 g 1.0359
5 g 1.0460
Peru MoO3 1.0130 1.0000
1 g 1.0380 1.0000
5 g 1.0380 1.0000

All factors are stated so that an uncorrected measured activation rate has to be multiplied with these to obtain the activation rate in a very small sample

TCC stands for true coincidence summing, ZDT denotes zero-dead time counting

Unity values indicate that the participant deemed the correction negligible, blank values that no data were provided by the participant