Table 1.
Prostate Restriction Spectrum Imaging protocol
Coil: Cardiac, no endorectal coil is required |
Repetition time: 9900 ms |
Echo time: 69–72 ms |
Field of View: 200–260 mm |
Matrix: 96×96 |
Slice thickness: 3.0 mm |
b values (s/mm2): |
0: alternate phase encoding direction for distortion correction |
125: 6 diffusion gradient directions |
375: 6 diffusion gradient directions |
1000: 15 diffusion gradient directions |