Table 2. Between-day imprecision of central laboratory automated analyzers (Lab).

Level 1 Level 2 Level 3 Average CV (%)
QC Mean CV (%) QC Mean CV (%) QC Mean CV (%)
Na+, mmol/L 122 122.8 0.7 ND ND ND 153 153.4 0.7 0.7
K+, mmol/L 3.93 3.96 0.7 ND ND ND 6.64 6.59 1.1 0.85
Cl-, mmol/L 89.9 90 0.65 ND ND ND 111 110.8 0.75 0.7
HCO3-, mmol/L 10 10.7 3.5 16 17.4 3.9 27 30 2.2 3.2
Glu, mg/dL 103 101 1.65 ND ND ND 240 240 1.35 1.5
Hb, g/dL 4.6 4.7 1.13 12 11.9 1.02 15.6 15.5 1.01 1.05
Hct, % 14.4 14.8 0.9 36.1 36 0.86 47.6 47.8 0.9 0.89

ND is Not Done, QC is quality control (target), CV is Coefficient of Variation, Na+ is sodium, K+ is potassium, Cl- is chloride, HCO3- is bicarbonate, Glu is glucose, Hb is hemoglobin, and Hct is hematocrit.