Parameter / Function |
Ancillary equipment |
Range of use |
Accuracy |
Technical Details |
Positioning |
Sample stage |
|
|
max. load 600 kg |
X |
0 - 360 mm |
0.05 mm |
Y |
0 - 330 mm |
0.05 mm |
Z (beam axis) |
manually, 600 mm |
|
θr (rotation) |
0° - 360° |
0.002° |
θt (cradle) |
± 30° |
0.002° |
Ambient temperature |
three-level multi-position Al holder (Cd masks) |
|
|
3x9=27 wide quartz cells |
3x12=36 narrow quartz cells |
three positions |
|
|
3 large quartz cells (diameter Φ = 5 cm) |
Al holder (B4C mask) |
Temperature |
thermostat (oil bath) + |
from -25 °C to 200 °C |
± 0.5 °C |
4 positions |
small copper block |
in air or in vacuum chamber (with sealed cuvettes) |
thermostat (water bath) + two-level multi-position |
from 5 °C to 85 °C |
± 0.2 °C |
2x9=18 wide quartz cells; 2x12=24 narrow quartz cells |
Al block |
thermostat + high precision oven |
from 10 °C to 120 °C |
< 0.1 °C |
1 position, wide quartz cell |
Peltier-thermostated cuvette-holder |
from -20 °C to 140 °C |
± 0.2 °C |
8 positions (all types of quartz or bras sandwich type cells cells); controlled atmosphere below 5 °C |
(B4C mask) |
Pressure |
HP cell SANS + thermostat (water bath) |
up to 5,000 bar |
|
Temperature in the range from 5 °C to 85 °C |
Low-temperature |
Cryostat with sapphire windows |
down to 50 K |
|
|
Rheometry |
Rheometer; steady state and oscillatory modes |
|
|
|
Humidity |
Humidity Cell |
5% to 95% |
|
Temperature in the range from 15 °C to 60 °C |
In-situ FT-IR20
|
FT-IR spectrometer |
|
|
Sample cells with ZnSe windows |