Skip to main content
. 2017 Jan 12;4(1):011008. doi: 10.1117/1.JMI.4.1.011008

Fig. 10.

Fig. 10

Cathode–cathode time difference histogram before and after correction for systematic errors. FWHM coincident time resolution of 37±5  ns was calculated by fitting a Gaussian curve to the coincident peak (circle points). The offset in the mean value of the histogram before calibration is due to the line delays intrinsic in data acquisition hardware.