Skip to main content
. 2016 Dec 2;31(1):53–61. doi: 10.1038/eye.2016.260

Table 1. Mean absolute difference and percentage error of neurosensory retinal tissue, foveal central subfield thickness, and total volume in different sampling groups.

Scan density Mean SD Range Scan density Mean SD Range
  Neurosensory retinal FCS thickness
Total neurosensory retinal volume
128 262.46 107.1 132.9–733.70 128 7.25 0.86 5.97–10.21      
64 262.56 106.9 133.7–732.1 64 7.25 0.86 5.98–10.23      
32 265.17 119.1 135.3–731.1 32 7.18 0.93 5.97–10.22      
16 263.33 102.9 147.5–727.1 16 7.25 0.85 5.96–10.2      
8 268.9 99.4 149.8–713.9 8 7.26 0.86 5.97–10.2      
4 266.57 96.3 119.3–696.9 4 7.25 0.87 5.97–10.42      
Absolute difference (μm)
Absolute difference (mm3)
     
64 1.21 1.05 0–4.10 64 0.01 0.01 0–0.04      
32 2.69 2.49 0.1–10.90 32 0.01 0.01 0–0.07      
16 5.97 6.9 0–34.50 16 0.02 0.02 0–0.06      
8 14.38 14.02 0.1–47.60 8 0.04 0.03 0–0.14      
4 25.64 27.68 0–112.40 4 0.1 0.08 0–0.30      
Percentage error (%)
Percentage error (%)
     
64 0.49 0.43 0–1.85 64 0.13 0.1 0–0.52      
32 1.11 0.38 0.03–4.90 32 0.2 0.21 0–0.91      
16 2.46 3.05 0–13.62 16 0.24 0.24 0–0.79      
8 6.3 6.95 0.05–26.14 8 0.56 0.5 0–2.3      
4 9.65 10.42 0–55.63 4 1.33 1.03 0–4.04      

Abbreviations: FCS, foveal central subfield; Scan density, number of B-scans.