Skip to main content
. 2017 Jan 17;12(1):e0170309. doi: 10.1371/journal.pone.0170309

Table 2. Comparison of the detection of vertebral metastases with and without CAD.

Reader A Reader B
CAD (–) CAD (+) CAD (–) CAD (+)
Number detected 47 57 40 44
Number of true positives 33 38 36 39
Number of false positives 14 19 4 5
Number of false negatives 13 8 10 7
Sensitivity 0.717 0.826 0.783 0.848
Specificity 0.981 0.974 0.995 0.993
AUC on ROC curve analysis 0.849* 0.902* 0.889 0.910

AUC, area under the curve; ROC, receiver operating characteristic; CAD, computer-aided diagnosis