Skip to main content
. 2017 Jan 4;2017:2149145. doi: 10.1155/2017/2149145

Table 2.

Summary of the intrasession repeatability outcomes for the corneal astigmatism and axis measurements obtained by means of the VX120 system depending on the magnitude of corneal astigmatism.

S w Corneal astigmatism ≤0.75 D Corneal astigmatism >0.75 D p value
K AST (D) 0.10 0.13 0.148
K ASTAX (°) 7.25 3.44 0.001
3KAST (D) 0.15 0.28 0.001
3KASTAX (°) 14.72 5.92 0.001
5KAST (D) 0.11 0.19 0.002
5KASTAX (°) 8.25 4.19 0.002
7KAST (D) 0.13 0.15 0.090
7KASTAX (°) 9.43 6.42 0.002

Mean keratometric astigmatism (KAST) and its axis (KASTAX), mean astigmatism (3KAST) and its axis (3KASTAX) in a 3 mm central zone, mean astigmatism (5KAST) and its axis (5KASTAX) in a 5 mm central zone, and mean astigmatism (7KAST) and its axis (7KASTAX) in a 7 mm central zone.