Skip to main content
. 2004 Nov;42(11):5199–5204. doi: 10.1128/JCM.42.11.5199-5204.2004

TABLE 5.

Performance of the target-capture PCR assay with the QCMD panel

QCMD code Subtype HBV concn
Accuracy (%)
Calculated by target-capture PCR
Published by QCMD (copies/ml)
IU/ml Copies/mla
HBV-01 A 547,239 1.4 × 106 1 × 106 140
HBV-02 A 288 0.74 × 103 1 × 103 74
HBV-03 0 0 0 100
HBV-04 D 14,265 0.37 × 105 1 × 105 37
HBV-05 A 4,077 1 × 104 1 × 104 100
HBV-06 A 1,155 3 × 103 3 × 103 100
HBV-07 A 87 226 200 113
HBV-08 A 39,445 1 × 105 1 × 105 100
a

Each result is the average for triplicate samples in IU per milliliter converted to copies per milliliter with the following formula: 1 IU = 2.6 copies.