Skip to main content
. 2016 May 3;89(1062):20150489. doi: 10.1259/bjr.20150489

Table 2.

Multileaf collimator errors and gamma pass rates following delivery of centre specific plans. Median values are also shown (minimum and maximum)

Parameter TrueBeam (N = 2) (n = 4) 2300IX (N = 5) (n = 10)
RMS values for leaves in Bank A (mm) 0.13 (0.11–0.14) 0.41 (0.24–0.46)
RMS values for leaves in Bank B (mm) 0.13 (0.12–0.14) 0.42 (0.25–0.47)
Gamma pass rate (3%3mm) 97.4 (84.0–100.0) 100.0 (98.9–100.0)
Gamma pass rate (2%2mm) 78.3 (50.5–97.9) 98.2 (85.4–99.7)

n, number of arcs analyzed; N, number of centres; RMS, root mean square.