Skip to main content
. 2016 Apr 14;89(1062):20160108. doi: 10.1259/bjr.20160108

Table 1.

Percentage change in signal-to-noise ratio (SNR) and percentage uniformity (U) values without and with the carbon fibre insert in place for the spin-echo (SE), turbo spin-echo (TSE) and gradient-echo (GE) sequences with the MultiTransmit switch (MX) turned “OFF” and then “ON”. The 32-element body coil configuration was used in this setup

RF transmission Normal couch
SNR change (%)
U (%)
SE TSE GE SE TSE GE
MX OFF



81
81
80
MX ON
+3
−3
+2
82
83
80
  Carbon fibre insert
SNR change (%)
U (%)
MX OFF −89 −82 −88 39 40 60
MX ON −77 −65 −83 54 55 61