. 2016 Apr 14;89(1062):20160108. doi: 10.1259/bjr.20160108

Table 2.

Percentage change in signal-to-noise ratio (SNR) and percentage uniformity (U) values without and with the carbon fibre insert in place for the spin-echo (SE), turbo spin-echo (TSE) and gradient-echo (GE) sequences with the MultiTransmit switch (MX) turned “OFF” and then “ON”. The scanner's quadrature body coil configuration was used in this setup

RF transmission Normal couch
SNR change (%)
U (%)
SE TSE GE SE TSE GE
MX OFF



96
96
96
MX ON
+29
−1.7
+0.87
97
96
96
  Carbon fibre insert
SNR change (%)
U (%)
MX OFF −80 −63 −82 74 74 86
MX ON −39 −12 −64 95 95 94