Table 5.

Multi-centre pQCT cross-calibration with EFP. As per ISCD recommendations, linear regression models determined slopes and intercepts related the European Forearm Phantom (EFP) bone mimic variables between the referent Hamilton pQCT scanner (model: XCT2000) and each of the scanners at other sites. 95% confidence intervals (CI) (lower, upper) for slope and intercepts were reported. Calibration data were obtained from the Canadian Multicentre Osteoporosis Bone Quality Study[22].

Bone variable Slope (95%CI) Intercept (95%CI) Slope (95%CI) Intercept (95%CI)
Vancouver XCT2000 Vancouver XCT3000
Tot.Ar (mm2) 1.02 (1.01,1.02) 5.90 (4.08,7.73) 1.02 (1.01,1.02) 5.31 (3.93,6.69)
vBMDi (mg/cm3) 0.97 (0.94,1.00) 23.16 (10.88,35.45) 0.99 (0.97,1.00) 13.85 (6.20,21.49)
Ct.Ar (mm2) 0.98 (0.89,1.06) 0.23 (-4.33,4.78) 1.00 (0.93,1.08) 1.04 (-2.76,4.83)
vBMDc (mg/cm3) 0.98 (0.89,1.07) 17.77 (-58.59,94.12) 0.92 (0.84,1.00) 69.70 (1.58,137.83)
Tb.Ar (mm2) 1.02 (1.01,1.02) 2.57 (1.72,3.43) 1.02 (1.01,1.02) 2.35 (1.75,2.94)
vBMDtr (mg/cm3) 1.17 (1.10,1.24) -27.17 (-44.85,-9.49) 1.19 (1.09,1.28) -31.15 (-53.34,-8.96)
Saskatoon XCT2000 Toronto XCT2000
Tot.Ar (mm2) 1.00 (1.00,1.00) -0.18 (-1.36,1.00) 1.00 (1.00,1.00) -0.75 (-1.76,0.25)
vBMDi (mg/cm3) 1.00 (0.99,1.01) -2.75 (-6.80,1.29) 1.01 (1.00,1.01) -8.21 (-10.65,-5.76)
Ct.Ar (mm2) 0.99 (0.92,1.06) 0.20 (-3.62,3.22) 0.97 (0.87,1.07) 0.65 (-4.57,5.86)
vBMDc (mg/cm3) 0.96 (0.83,1.09) 27.41(-80.22,135.04) 0.91 (0.78,1.05) 67.96 (-43.67,179.59)
Tb.Ar (mm2) 1.00 (1.00,1.00) -0.09 (-0.60,0.42) 1.00 (1.00,1.00) -0.40 (-0.85,0.05)
vBMDtr (mg/cm3) 0.99 (0.97,1.01)
Kingston XCT2000
Tot.Ar (mm2) 1.00 (1.00,1.00) -0.57 (-1.45,0.32) - -
vBMDi (mg/cm3) 0.98 (0.97,0.99) -3.92 (-7.39,-0.44) - -
Ct.Ar (mm2) 0.93 (0.84,1.02) 0.62 (-4.32,5.56) - -
vBMDc (mg/cm3) 0.92 (0.81,1.02) 50.90 (-38.96,140.75) - -
Tb.Ar (mm2) 1.00 (1.00,1.00) -0.23 (-0.65,0.20) - -
vBMDtr (mg/cm3) 0.97 (0.96,0.98) -0.91 (-3.84,2.01) - -