Skip to main content
. 2016 Dec 28;15(Suppl 2):124. doi: 10.1186/s12938-016-0251-5

Table 2.

The comparison of lesion size determined by the 50 °C isotherm and the CEM43 °C at t = 60 s

Voltage (V) Lesion size of 50 °C (mm2) Lesion size of CEM43 (mm2) Overestimate rate (%)
30 9.4487 7.0958 24.9
35 14.084 11.516 18.23
40 19.091 15.619 18.18
50 28.417 23.637 16.82
60 35.568 31.168 12.37