Table 1.

Hemodynamic parameters of the parent vessel in the three different methods

Average of velocity inside aneurismal sac (m/s) Area average of WSS on the aneurysm (Pa) Area average of WSS on the parent vessel (Pa)
Realistic (X) 0.0829 0.8317 4.1118
FVS (Y) 0.0908 0.8212 3.9075
FEM (Z) 0.0905 0.8155 4.2283

X corresponding hemodynamic parameters in realistic phantom deployment of Enterprise stent; Y corresponding hemodynamic parameters in FVS simulation of Enterprise stent; Z corresponding hemodynamic parameters in FEM simulation of Enterprise stent; WSS wall shear stress