Skip to main content
. Author manuscript; available in PMC: 2017 Jan 24.
Published in final edited form as: Nat Protoc. 2016 Oct 27;11(12):2328–2356. doi: 10.1038/nprot.2016.114

Table 1.

Troubleshooting table.

step problem possible reason solution
4 Low yield of
alcohol 2
Ethylenediamine is old
Reaction interfered with by
O2 and moisture
Product destroyed during
reaction quenching
Use either re-distilled ethylenediamine or fresh out of the bottle
Make sure that the reaction is protected under dry N2 or Ar

The reaction quenching is exothermic; add the quencher (1 N HCl)
slowly and make sure that the flask is well cooled
7 Low yield of
aldehyde 3
Aldehyde 3 is volatile Avoid excess heating (above 35 °C)of the rotary evaporator
water bath
11A(iv) Low yield of
HNE(alkyne)
HNE(alkyne) is not stable It is prone to air oxidation and polymerization. Carry on nonstop
through steps involving reaction workup, rotary evaporator,
chromatography, to final storage
11B(viii) Product 7 partially
deprotected
One of the isomers of
product 7 is not stable
Product 8 (deprotected 7) can be separated from 7 by
chromatography on silica gel, or it can be carried through to
the next step
30 Low yield of 14 Reaction yield can be reduced by adventitious
O2 and/or moisture
Make sure that all the solvents are dry and that the reaction
is protected under dry N2 or Ar. Thoroughly degas the
solvents used
31 Incomplete
deprotection
of 14
Some of the catalyst
(10% Pd/C) may be
deactivated
Add more catalyst (10% Pd/C) and extend the reaction time.
Typically, high catalyst loading can be tolerated. Ensure that the
reaction is tightly sealed and that an atmosphere of hydrogen is maintained throughout
36 Difficult to
remove impurities
from HtPHA
Polarity of impurities is
close to that of HtPHA
Run flash chromatography with gradient eluent, from 1:3 to 1:5
(vol/vol, hexanes: EtOAc)
Final product
decomposition
Light-induced photo-
uncaging of the final
product
During the synthesis (especially from 14 to 15, and 15 to HtPHA),
protect the reaction flask from stray light
40A(x) No Cy5 signal on gel Failure of click coupling Check lysate protein concentration. The concentration should be
around 1.0 mg/ml
Check each reagent in the click coupling step. Make sure that
all of them are freshly prepared and that the concentrations are
correct. Mix the reaction well
Use cells treated globally with HNE(alkyne) as a positive control
High Cy5 background Old SDS-PAGE running buffer Use fresh SDS-PAGE running buffer
Rinse the gel several times and analyze the gel after each rinse to
obtain the optimal result with highest signal-to-noise ratio
Let the dye font run out completely before imaging the gel
Incomplete TEV
cleavage
Loss of TEV activity Use a fresh aliquot. Avoid multiple freeze–thaw cycles
Mix well after adding His6-TEV-S219V
Increase TEV amount and incubation time for TEV cleavage
40B(xxviii) Low yield of
pulldown protein
Affinity protein purification
condition is not optimal
Protein is not eluting or
eluting prematurely



Protein is unstable after
modification103
This protocol is optimized for His6-Halo-TEV-Keap1. Further
optimization may be required for other proteins
Monitor the protein in washes through SDS–PAGE. If premature
elution is observed, decrease the concentration of imidazole in
the wash buffers. Conversely, if no elution is observed, increase
imidazole concentration to 200–400 mM or elute with Laemelli
buffer to validate binding
Add a proteasome inhibitor (bortezomib) in the lysis buffer
and/or to cells
Modification
not found
Targeting is not efficient.

Modification is reduced
Inefficient MS conditions
Check modification of protein by Cy5 labeling, as described
in Step 3A
Make sure that TCEP is used instead of DTT in the sample preparation
Adjust the ionization temperature to obtain optimal results.
A low ionization temperature may lead to poor ionization and
decrease sensitivity. A high ionization temperature may cause
loss of modification
40C(xi);
40D(viii)
Low Firefly/Renilla/
GFP signal intensity
Significant loss of cells
during rinsing




Excessive cell death


Low transfection efficiency

Instrument setting is
not optimal
Response timing is
not optimal
Perform the rinsing with care. Add the medium slowly along
the side wall of the culture dish. Adding and removing medium
should be done as gently as possible but also as efficiently
as possible: wash plates sequentially (2 or 3 plates at a time)
to avoid drying out. Mark the position on the plate where cells
are washed
Use the suggested cell density for transfection. If using a
different transfection reagent or different cells, optimize
transfection conditions
Optimize transfection conditions for the cell type and the
reagent used
Run a positive control. For flow cytometry, cells transfected
with GFP can be used. Adjust gain (i.e., laser power) if
necessary
Set up a time course for measurements (recommended time
points for pilot trials: 4 h, 12 h, 18 h post T-REX light
exposure)
Batch-to-batch
variability in targeting/ARE
results
Difference in experimental
setup and execution
Count cells and seed the numbers as specified in the protocol.
Transfect and perform experiments at similar confluence
Use cells at lower passage number (lower than 6–7 continuous
passages)
Activation of
ARE-luciferase with
light alone or with
pre-HNE alone
Release of pre-HNE due
to stray light
Stressed cells
Protect photocaged-precursor-treated samples from stray light

Count cells before seeding. A cell density that is too low or
too high can stress the cells, leading to higher background