Table 2. Determination of DA in human serum and urine samples by using GA-RGO/AuNPs modified electrode.
Sample | Spiked (μM) | Found (μM) | Recovery (%) | RSD (%) |
---|---|---|---|---|
Human serum | 5.0 | 4.76 | 95.2 | 2.4 |
5.0 | 4.89 | 97.8 | 2.7 | |
Urine sample | 5.0 | 4.84 | 96.8 | 3.1 |
5.0 | 4.93 | 98.6 | 3.3 |
The relative standard deviation (RSD) is related for n = 3.