TABLE 3.

Analytical performances of the BYG Carba test, β Carba test, Rapidec Carba NP test, and Neo-Rapid Carb screen kit

Parametera Results forb:
BYG BCT RCNP NRCK
Enterobacteriaceae (n = 198)
    Sensitivity (%) 100 97.3 91.9 89.2
    95% CI 100–100 93.4–100 85.4–98.4 81.8–96.6
    Specificity (%) 98.9 97.7 83.9 89.7
    95% CI 96–100 93.7–100 74.1–93.8 81.5–97.8
    PPV (%) 99.1 98.2 87.9 91.7
    95% CI 96.9–100 95–100 80.4–95.5 85–98.3
    NPV (%) 100 96.6 89 86.7
    95% CI 100–100 91.8–100 80.4–97.7 77.7–95.6
Pseudomonas spp. (n = 89)
    Sensitivity (%) 87.3 90.9 90.9 92.7
    95% CI 76–98.5 81.2–100 81.2–100 84–100
    Specificity (%) 94.1 94.1 88.2 88.2
    95% CI 84–100 84–100 74.4–100 74.4–100
Acinetobacter spp. (n = 41)
    Sensitivity (%) 72.7 75.8 36.4 27.3
    95% CI 53.3–92.1 57.1–94.4 15.4–57.3 7.9–46.7
    Specificity (%) 100 87.5 75 100
    95% CI 100–100 58.3–100 36.7–100 100–100
a

CI, confidence interval; PPV, positive predictive value; NPV, negative predictive value.