TABLE 4.
Technical characteristics of BYG Carba test, β Carba test, Rapidec Carba NP test, and Neo-Rapid Carb screen kit
Characteristic | Results fora: |
|||
---|---|---|---|---|
BYG | BCT | RCNP | NRCK | |
Validated group of isolates (or recommended by manufacturer) | Enterobacteriaceae | Enterobacteriaceae | Enterobacteriaceae | Enterobacteriaceae |
P. aeruginosa | P. aeruginosa | P. aeruginosa | ||
A. baumannii | ||||
Maximal incubation time (min) | 60 | 30 | 160 | 90 |
Total hands-on time (min) | 5 | 2 | 4 | 6 |
No. of handling steps | 1 | 1 | 3 | 2 |
Ease of use | Easy, little training required | Very easy, no training required | Easy, little training required | Easy, little training required |
Catalog price per unit test in Belgium | Not commercially available | 6.6 USDb | 8.8 USD | 2.4 USD |
Positive criteria | Signal attaining a fixed cutoff | Orange, red, purple | Orange-red to yellow, with a red negative control | Orange to yellow, with a red negative control |
Overall doubtful results (%)c | 0 | 3 | 9.1 | 4.9 |
Overall noninterpretable resultsc | NRd | NR | 0.3 | 3.4 |
Overall reproducibility (%)e | 95 | 85 | 100 | 90 |
BYG, BYG Carba test, BCT, β Carba test, RCNP, Rapidec Carba NP test, NRCK, Neo-Rapid CARB screen kit.
USD, U.S. dollars.
Proportion of doubtful and noninterpretable results among all tested isolates.
NR, not reportable as no internal negative control is provided with the test.
Proportion of isolates showing full agreements when tested in triplicate in a reproducibility evaluation (total n = 20).