Skip to main content
. 2017 Jan 25;55(2):545–551. doi: 10.1128/JCM.02161-16

TABLE 1.

Optimized cycles and parameters used in this study for adequate sterilization of simulated waste

Parameter PPE and dry trash Saturated linen Liquid
Cycle type Vacuum Vacuum Liquid
Sterilization time (min) 30 60 120
Sterilization temperature (°C) 134 134 123
Dry/cool time (min) 1 10 35
Purge time (min) 4 5 14
Precharge (psig) 20 20 20
No. of prevacsa 3 3 3
Prevac vacuum point (in. Hg) 10 10 10
Autoclave bag type and configuration Standard clear bags, double bagged
Bag closure Clamped outer bag, loosely twisted inner bagb
a

Prevac, prevacuum pulse.

b

Remove clamp and place in tray prior to autoclaving.